


Next:Spatial
Variations in Quantum
Up:Spatially-averaged
Relative Quantum Efficiency
Previous:Uncertainties
in the Cross-calibration
Internal Consistency of Relative
Quantum Efficiency Measurements made at MIT CSR.
As noted above, the flight detector I0 (w203c4r) was calibrated with respect
to both w190c3 and w103c4. Since we also cross-calibrated w190c3 and w103c4
with respect to one another, it is possible to check the consistency of
these measurements. This comparison is shown in Table 4.56.
Excluding the point at 8 keV (which was used in the reference detector
cross-calibration), the mean difference is
,
consistent with zero, where we have taken the standard deviation of the
mean as the error. The RMS about the mean is 0.0085. Assuming each of the
3 relative QE measurement sets has the same error, and that the error is
independent of energy, then, to a good approximation, the error in any
single measurement is
.
Table 4.56: Reproducibility of Relative Quantum Efficiency
Measurement of Detector I0
| Flight |
Position |
Mean Relative Efficiency vs. Energy (keV) |
Reference |
| Device |
|
0.525 |
0.677 |
1.740 |
2.015 |
4.509 |
5.894 |
8.040 |
Device |
| |
|
O |
F |
Si |
P |
Ti |
Fe55 |
Cu |
|
| w203c4r |
I0 |
0.972 |
0.992 |
0.996 |
 |
 |
0.977 |
 |
w190c3 (meas.) |
| w190c3 |
Ref. |
0.890 |
0.767 |
0.986 |
 |
 |
1.058 |
 |
w103c4 (meas.) |
| w203c4r |
I0 |
0.866 |
0.760 |
0.982 |
 |
 |
1.034 |
 |
w103c4 (calc.) |
| w203c4r |
I0 |
0.848 |
0.756 |
0.977 |
 |
 |
1.030 |
 |
w103c4 (meas.) |
| Difference |
line 3 - line 4 |
0.018 |
0.004 |
0.005 |
-0.008 |
-0.001 |
0.004 |
 |
|
Cross-calibration
of w190c3 vs. w103c4 uses w203c4r data; see text |
| Mean Difference, excluding 8 keV point : 0.0037 |
| RMS Difference, excluding 8 keV point : 0.0085 |
|
This level of error, while probably acceptable for ACIS calibration
purposes, is one order of magnitude larger than the limiting uncertainty
imposed by photon counting statistics. A number of factors may contribute
to this excess error. The two reference detectors used are known to have
slightly different pileup characteristics, but the same pileup model was
used in the analysis of both data sets. One reference detector (w103c4)
was operated with an older (pre-ACIS) generation of electronics that is
somewhat noisier than the ACIS electronics; the slight differences in response
function may contribute some error. Finally, relative misalignment between
flight and reference CCDs can influence low energy ratios (C, O, F) due
to the non-uniformity of these sources. (Some misalignment on order 50
pixels seems evident in a few cases, despite the apparent precision achieved
with the alignment system.) This is less important for higher energies,
which make use of a commercial X-ray tube and have a flatter illumination
pattern at the CCD location (primarily due to a larger source-detector
separation).
Systematic errors in the modelling of the relative calibration measurements
are discussed in detail in Section 4.8.



Next:Spatial
Variations in Quantum
Up:Spatially-averaged
Relative Quantum Efficiency
Previous:Uncertainties
in the Cross-calibration
Please address comments and questions to Dr. John Nousek ( nousek@astro.psu.edu
)