Table 4.67 lists the
mean gains (all four quadrants of a chip) measured at XRCF-FF and ISIM-TV1.
Taking the XRCF-FF values as a baseline, the I-Array chips (imaging CCD's
and S2 and S3) have higher gains at ISIM-TV1, while the outer S-Array chips
(S0,S1,S4,S5) have approximately the same gains. The gain is known to be
dependent on DEA temperature and focal plane temperature. As the effect
of the focal plane temperature on the gain is an order of magnitude smaller
than that of the electronics temperature, we will focus on the later as
a possible explanation of the difference in gains between the two measurements.
The mean temperature dependence relationship is: d(ln Gain)/dT=
10-4
(see Section 4.9.2 for details).
The thermal conditions for both calibrations were nominally the same: focal
plane temperature of -110
C, and DEA -Z plane temperature of +13
C. Allowing for a drift in the DEA temperature of
C , we would expect a gain variation on order
0.002, much higher than that observed for the I-array chips. Figure 4.98
plots d(ln Gain) for each quadrant of all ten chips.
|
Systematic variations, order of a few degrees, are expected in the
DEA temperature; the dashed lines indicate the values of d(ln Gain) expected
for
T=
5
C. The outer S-Array chips are mostly distributed between this range, indicating
that the gains measured with the ExtCalSrc for these chips are stable,
within systematic uncertainties. Another dashed line indicates the value
of d(ln Gain) for
T=25
C; the I-Array chips measured with the ExtCalSrc are tightly grouped abou
this line. At this time, the source of the gain shift is not known. However,
given the uncertainties in the operating conditions and the thermal environment
during the ISIM-TV tests, there exists the distinct possibility that a
strong thermal gradient existed inside the DEA unit. This condition can
explain the apparantly paradoxical behavior that some of the gains indicate
that the DEA -Z plane was in fact
13
C, while the other chips indicate that the the electronics were
25
C hotter. This scenario requires that the boards used to read the I-Array
chips were in the hottest part of the gradient and furthest from the DEA
-Z plane temperature sensor. Further evidence for this hypothesis comes
from Internal Calibration Monitor (ICM) data. The ICM can also be used
to determine gains, but only for four quadrants. The ICM data was also
taken as part of the XRCF-FF tests, as well as during the second ISIM thermal
vacuum test, ISIM-TV24.8.
The XRCF-ICM gains (blue symbols in the plot) are also distributed near
the
5
C lines, establishing that the ICM source can provide an accurate gain
measurement. During ISIM-TV2, however, the ICM gains (orange symbols in
the plot) are also grouped about the 25
C line. The fact that the two ISIM thermal vacuum tests were separated
by more than two months indicates that whatever condition caused
the shift in the ExtCalSrc gains was also present during the subsequent
testing. If the mechanism was in fact a thermal gradient, it must have
have been stable.
Figure 4.98: The difference in gain measured at XRCF-FF and ISIM-TV1. The figure plots d(ln Gain)=Delta Gain/Gain, where the gain from XRCF-FF is taken as a base line. The red points compare the gains calculated from the ExtCalSrc data at ISIM-TV1, the blue points compare the gains calculated fro m the ICM source at XRCF-FF, and the orange points compare the gains calculated from t he ICM source at ISIM-TV2. The dashed lines and labels indicate the change in DEA temp erature required to shift the gain by that amount. The ICM source only allows gain meas urements for S2 Quad D and S3 Quads A-C.
