


Next:Search
for Effects of
Up:ACIS
CCD Sensitivity to
Previous:Ionizing
Dose Effects
Effects of Energetic Protons
High-energy proton dose tests were made at the Harvard Cyclotron Laboratory
where an engineering front illuminated CCID-17 was exposed to a well-characterized
proton beam with 40 MeV peak energy. Exposures were made at room temperature
in ambient air with a lead aperture that allowed irradiation of a controlled
section of the image array. Irradiation occurred in incremental steps to
a final total of 400 Rads. A minimum 48 hour period at ambient condition
elapsed between exposure and all subsequent characterization. While there
is evidence to suggest changes in post radiation behavior immediately following
exposure with time scales of hours no changes were seen following the 2
day period - all damage features appear permanent and unaffected by subsequent
temperature cycles.
As with the X-ray test the most noticeable result of proton beam exposure
was the production of an area of enhanced dark current coincident with
the exposed region. This dark current could not be measured for temperatures
below -90 c. No flickering or hot pixels were detected at any temperature.
Proton damage is expected to produce traps whose presence can be seen
by an increase in CTI, and indeed a gradual increase in CTI at a rate of
2 x 10-5 per kRad was seen when tested with a spatially uniform
25 count/sec/output Fe-55 source. A clocking method was used to create
an adjustable line of charge in the image array. At -120 c this sacrificial
charge technique only influenced pixels in the next few rows suggesting
a detrapping time of order 100 microseconds. At -90 c the CTI is four times
greater and the utility of the charge injection method more pronounced
although improvements of any performance measure such as CTI or resolution
could never be increased beyond levels achieved by normal operation at
-120 c. Additional proton irradiation of just the frame store area had
no affect upon the measured CTI but did continue to degrade the gain and
resolution of the device. This is a consequence of the way in which CTI
is measured.
Figure 4.121 shows the
observed variation of CTI with proton dose.
Figure 4.120: Dark current map
showing circular area of exposure to 5.9 keV photons. Total dose was 900
rads (Si); this map was obtained with the device at -60C. At nominal operating
temperature (-120C) no increase in dark current is measurable. Device quadrants
0-3 are from left to right; the serial register is toward the top of the
picture.
Figure 4.121: Charge transfer
inefficiency (CTI) vs. accumulated ionizing dose of 40 MeV protons for
a front-illuminated ACIS CCD. The curves labelled c0,c1, and c2 pertain
to exposed areas of the device. The curve labelled c3 pertains to an area
well-shielded from proton radiation.



Next:Search
for Effects of
Up:ACIS
CCD Sensitivity to
Previous:Ionizing
Dose Effects
Please address comments and questions to Dr. John Nousek ( nousek@astro.psu.edu
)